0.38mps Biological Safety Cabinet Class 2 LCD display
Product Details:
Place of Origin: | CHINA |
Brand Name: | HAIYING |
Certification: | CE |
Model Number: | BSC-1000A2 |
Payment & Shipping Terms:
Minimum Order Quantity: | 1 set |
---|---|
Price: | Negotiable |
Packaging Details: | Wooden case |
Payment Terms: | L/C, D/A, D/P, T/T, Western Union, MoneyGram |
Detail Information |
|||
Type: | Blood Testing Equipments | Model: | BSC-1000A2 |
---|---|---|---|
Overall Size: | 1200×800×2120MM | Purification Area Size: | 1000×700×600MM |
Product Name: | Biosafety Cabinet | Inside Of The Door: | 0.38±0.025M/S |
Front Inflow Air: | 0.48±0.025M/S | Vibration/half Peak Amplitude: | ≤3μm(X·Y·Z) |
Highlight: | 0.38mps biological safety cabinet class 2,0.38mps safety cabinet class 2,0.3um biological safety cabinet class 2 |
Product Description
For chemical laboratories Federal guarantee Class II A2 Biosafety Cabinets
Class II A2 Biosafety Cabinets Product introduction:
It is widely used in biosafety protection and isolation equipment in microbiology, biomedicine, biological laboratories and other laboratories. It adopts advanced air purification technology and negative pressure cabinet design to realize the protection of the environment, personnel and samples and prevent harmfulness. The diffusion of suspended particles and aerosols.
BSC-II-A2 biological safety cabinet series is a level II A2 biological safety cabinet with 70% gas circulation and 30% gas exhaust.
Class II A2 Biosafety Cabinets product description:
◆ Humanized structural design;
◆ High-brightness LCD display, all operating status is displayed, clear at a glance;
◆ Excellent fan selection and strict air tightness test;
◆ Alarm system;
◆ Safe interlock protection design: interlock protection is set for misoperation, even if misoperation, it will not cause harm;
◆ The cabinet body and the bracket can be separated, and the height of the bracket can be customized and modified according to the actual situation;
◆ Safe and reliable performance guarantee; remote control reduces the direct contact between the user and the safety cabinet, and can protect the user more. It has an appointment timing function, can automatically disinfect, automatically switch the machine, and improve work efficiency.
Model | BSC-1000A2 |
Clean Grade | HEPA:ISO 5 calss ( Class100 ) |
ULPA:ISO 4 Class ( Class10 ) | |
HEPA Filter | HEPA:≥99.995% , @0.3μm |
ULPA:≥99.999% , @0.12μm | |
Air tightness | 99.99≤E≤99.999 |
Overall Dimensions | 1200×800×2120 mm |
Working Space | 1000×700×600 mm |
Inside of the door | 0.38±0.025m/s |
Front inflow air | 0.48±0.025m/s |
Installation of Class II A2 Biosafety Cabinets :
1. The biological safety cabinet shall not be placed on its side, impacted, or collided during transportation, and shall not be directly attacked by rain, snow, or sun exposure.
2. The working environment of the biological safety cabinet is 10~30℃ and the relative humidity is <75%.
3. The equipment should be installed on a horizontal surface that cannot be moved.
4. The equipment must be installed close to a fixed power socket. When the external exhaust system is not connected, the top of the device must be at least 200mm away from the obstacle at the top of the room, and the distance between the back and the wall must be at least 300mm to facilitate the smooth flow of external exhaust. Maintenance of safety cabinets.
5. In order to prevent airflow interference, it is required not to install the equipment in the passage of people, and not to make the operating port of the sliding front window of the biological safety cabinet face the door and window of the laboratory or be too close to the door and window. Places that may disturb the airflow.
6. When used in high altitude areas, the wind speed must be recalibrated after installation.