Detail Information

Type: Blood Testing Equipments Model: BHC-1000B2
Overall Size: 1500×750×1950MM Purification Area Size: 1300×650×580MM
Product Name: Biosafety Cabinet Inside Of The Door: 0.38±0.025m/s
Front Inflow Air: 0.48±0.025m/s Use: For Biological Laboratories
High Light:

Blood Class II B2 Biosafety Cabinet

,

lab biosafety cabinet class 2 b2

,

0.38mps Class II B2 Biosafety Cabinet

Product Description

For biological laboratories 100 level protection Class II B2 Biosafety Cabinet

Class II B2 Biosafety Cabinet Features:

1. Stainless steel operating area. Corrosion resistant, easy to clean, and equipped with a power socket.

2. LED lighting system. The lighting system design, equipped with LED tubes, can effectively disinfect the products in the work area.

3. Spring-loaded sliding door up and down. All windows can be opened and closed, and can be stopped at any height, so as to facilitate the entry and exit of the experimental equipment.

4. Electric front window.

5. Automatic wind speed, with filter block.

6. The side and back walls are made of a single piece of stainless steel.

7. Interlocking function: ultraviolet lamp and front window; ultraviolet lamp and blower, fluorescent lamp; blower and front window.

8. Lean forward 10°, which can provide the operator with a sense of comfort when working for a long time, reduce glare and extend into the work area to the greatest extent.

 

Model BHC-1000B2
Clean Grade HEPA:ISO 5 calss ( Class100 )
ULPA:ISO 4 Class ( Class10 )
HEPA Filter HEPA:≥99.995% , @0.3μm
ULPA:≥99.999% , @0.12μm
Air tightness 99.99≤E≤99.999
Overall Dimensions 1500×750×1950
MM
Working Space 1300×650×580
MM
Inside of the door 0.38±0.025m/s
Front inflow air 0.48±0.025m/s

 

Maintenance of Class II B2 Biosafety Cabinet:

In order to ensure the safety of the biological safety cabinet, the safety cabinet should be maintained and maintained regularly:

1.  The working area of ​​the safety cabinet should be cleaned and disinfected before and after each use.

2.  After the service life of the HEPA filter expires, it should be replaced by a professional who has received special training in biological safety cabinets.

3.  The laboratory biosafety manual promulgated by WHO, the U.S. biosafety cabinet standard NSF49 and the China Food and Drug Administration biosafety cabinet standard YY0569 all require one of the following conditions, the biosafety cabinet should be safety tested: installation is completed and put into use Before; annual routine inspection; when the safety cabinet is shifted; after replacing the HEPA filter and repairing internal components.

Safety inspection includes the following aspects:

1.  Inlet flow direction and wind speed detection: Inlet flow direction is detected on the working section by smoke method or wire method. The detection position includes the periphery and middle area of ​​the working window; the wind speed of the intake flow is measured by an anemometer.

2.  Detecting the wind speed and uniformity of the sinking airflow: Use an anemometer to measure the wind speed of the cross section evenly.

3.  Work area cleanliness detection: use dust particle timer to detect in the work area.

4.  Noise detection: The horizontal center of the front panel of the biological safety cabinet is 300mm outwards and 380mm above the work surface to measure the noise with the sound level.

5.  Illumination detection: set a measuring point every 30cm along the center line of the length of the worktable.

6.  Leakage detection of the cabinet: Seal the safety cabinet and pressurize it to 500Pa. After 30 minutes, connect a pressure gauge or pressure sensor system in the test area to detect with the pressure attenuation method, or detect with the soap bubble method.

 