Stainless Steel Low Speed Centrifuge Centrifugal Cavity Max 4000 R/Min
Product Details:
Place of Origin: | China |
Brand Name: | HAIYING |
Model Number: | TD4 |
Payment & Shipping Terms:
Minimum Order Quantity: | 1 Set |
---|---|
Price: | Negotiable |
Packaging Details: | Carton |
Payment Terms: | L/C, D/A, D/P, T/T |
Detail Information |
|||
Maximum Centrifugal Force (×g): | 2600 ×g | Maximum Speed(r/min): | 4000 R/min |
---|---|---|---|
Noise: | ≤60dB | Early Warning Function: | Overspeed、overtemperature、imbalance、etc |
Weight: | 18000 G | Dimensions (length×width×height): | 38×26×31 Cm |
Highlight: | Stainless Steel Low Speed Centrifuge,4000 R/Min Low Speed Centrifuge,4000 R/Min stainless steel centrifuge |
Product Description
Host technical parameters | |
product name | Stainless steel centrifugal cavity Max 4000 r/min TD4 Low Speed Centrifuge |
Maximum speed (r/min) | 4000 |
Maximum centrifugal force (×g) | 2600 |
Speed accuracy | ±50r/min (can be customized according to requirements, multiples of 10) |
Timing range | 1-99h59min/continuous/short time centrifugation |
Dimensions (length×width×height) | 380×260×310 mm |
TD4 Low Speed Centrifuge Common Failure Analysis:
1. After the power is turned on, the control panel displays normally. When the start button is pressed, the centrifuge works normally. But when the stop switch is pressed, the centrifuge has no response and cannot stop working. According to the analysis of the failure phenomenon, the failure should be in the stop control part of the control circuit part, which includes the stop switch, the power control board and the timer control board. Turn on the centrifuge, use a multimeter to check the stop switch, it is normal; after starting, measure the signal output from the power control board to the timer control board is AC 16V is normal, the fault should be on the timer control board, the timer control board can not execute the stop when the stop signal is obtained task. After troubleshooting one by one, it was found that the C9013 transistor on the timer control board and the capacitor with the model of 16V, 470μF broke down, and the new transistor and capacitor were worsened. The fault was eliminated.
2. After the power is turned on, the control panel displays normally. When the start button is pressed, the centrifuge works normally. But when the stop switch is pressed, the centrifuge cannot stop working immediately, but slowly stops after spinning for a long time. According to the analysis of the fault phenomenon, the fault should be in the stop brake control part of the control circuit part. The brake control part includes the model: WJ176-12V intermediate relay, the brake resistance R: 20-30Ω, the model: KBPC50A10M silicon stack and wire . Turn on the centrifuge and use a multimeter to measure the brake resistance, intermediate relays and wires. They are all normal. After turning on the machine, press the stop button and use the multimeter to measure the 4 corners of the silicon stack. There is AC 12V input but no DC 12V output. It is judged that the silicon stack is broken and replaced. New, troubleshooting.